Corresponding author: Hibba Dar, hibba.qayoom@gmail.com
DOI: 10.31662/jmaj.2025-0309
Received: July 1, 2025
Accepted: October 17, 2025
Advance Publication: December 5, 2025
Published: January 15, 2026
Cite this article as:
Dar H, Tabassum R, Rather Y, Farhat S. Qualitative and Quantitative Estimation of Ceftriaxone in Pharmaceutical Dosage Forms Using Reverse Phase High Performance Liquid Chromatography. JMA J. 2026;9(1):292-301.
Introduction: As the “pharmacy of the world,” India plays a critical role in global pharmaceutical supply. However, increasing reports of substandard and falsified medicines raise serious concerns. With Jammu and Kashmir witnessing repeated quality failures of ceftriaxone, this study aims to verify its quality using a refined high-performance liquid chromatography (HPLC)-based analytical method.
Methods: The ceftriaxone assay was performed using Agilent 1260 Infinity HPLC System with a C18 column and EZChromS1 software. A solvent system of acetonitrile and water (20:80, v/v) was used, with flow rate of 0.5 mL/min and ultraviolet detection at 254 nm. Standard solutions (20-70 μg/mL) were prepared from a 1,000 μg/mL stock for calibration. Nineteen injectable ceftriaxone brands were procured and coded for anonymity. Sample solutions were prepared at 50 μg/mL. The method was validated as per International Council for Harmonisation (ICH) guidelines for accuracy, precision, linearity, specificity, limit of detection (LOD), limit of quantification (LOQ), and robustness. Solution stability was assessed over 7 days at 2°C-8°C.
Results: The developed isocratic reversed-phase HPLC method showed good linearity (R2 = 0.9991) over the range 20-70 μg/mL, with LOD and LOQ of 5.88 μg/mL and 17.83 μg/mL, respectively. Intra-day and inter-day precision and accuracy showed percentage relative standard deviation (%RSD) values <2% and mean recovery within acceptable limits (96.9%-102.6%). Robustness was confirmed through variations in flow rate, mobile phase composition, detection wavelength, and column temperature, all yielding %RSD <2%. Analysis of 19 marketed formulations revealed that 89.5% complied with pharmacopoeial standards (90.0%-115.0% recovery), while 10.5% (CEF A17, CEF A19) were substandard, indicating significant quality variation among brands (p < 0.05).
Conclusions: A simple, sensitive, and cost-effective HPLC method was developed and validated for estimation of ceftriaxone in pharmaceutical dosage forms. Among 19 tested formulations, 89.5% complied with Indian Pharmacopoeial standards, while two formulations were found substandard, indicating variability in product quality.
Key words: ceftriaxone, RP-HPLC, method validation, drug quality assessment, substandard formulations
The aim of universal health coverage is that all people have access to the health services they need, when and where they need them, without financial hardship (1). To achieve this universal health coverage, the United Nations declared access to “safe, effective, quality, and affordable essential medicines and vaccines for all” as one of the Sustainable Development Goals (3.8) in their 2030 Agenda (2). Assuring the quality of medicines is crucial for providing safe and effective health care and reducing overall health care costs.
In terms of the pharmaceutical global market, India accounts for a significant share and has come to be known as the pharmacy of the world (3). By volume, India now ranks third worldwide when it comes to pharmaceutical production. The Indian pharmaceutical sector supplies over 50% of the global demand for various vaccines, 40% of the generic demand for the United States and 25% of all medicines for United Kingdom (4). Keeping in mind India’s growing position in the global pharmaceutical market, it becomes all the more imperative to keep a check on the quality of medicines being manufactured within the Indian pharmaceutical industry.
Molecules that are more commonly used, including antibiotics, have an increased likelihood of being counterfeited. In fact, among all classes of medicines, it was observed that quality issues in antimicrobials were reported 8-10 times more frequently (5). According to a World Health Organization (WHO) report released in 2017, 10.5% of medicines sold in low- and middle-income countries (LMICs), including India, are substandard and falsified (6). Beta-lactam antibiotics, especially penicillin and cephalosporins, have emerged as the most in-demand class of antimicrobials, with worldwide sales of approximately 65% of the total market of antibiotics (7). Beta-lactams also account for 50% of all reported cases of counterfeit antibiotics (8).
Among the substandard and falsified products reported to the WHO Global Surveillance and Monitoring System between 2013 and 2017, 17% are antibiotics (9). The prevalence of substandard antimicrobial agents varies from 0% to >50% globally (5). India has emerged as the top-most country in producing poor-quality antimicrobials (5), (8).
For assay and impurity profiling methods used in routine regulatory and quality control procedures, liquid chromatography―mainly high-performance liquid chromatography (HPLC) coupled with ultraviolet (UV) or fluorescence detection, has been established as the gold standard (10), (11), (12), (13).
In a study conducted by Shah et al. (14), 2013, a C18 column of 250 mm length and 4.6 mm diameter, with isocratic solvent delivery system was used. They employed a ternary mixture of acetonitrile (ACN), methanol, and triethyl amine buffer (pH 7) in ratio of 1:1:2 v/v as mobile phase, and analyzed the eluent at a wavelength of 240 nm, with a flow rate of 0.9 ml/min and a run time of 10 minutes. Gurupadayya and Disha (15), 2013, in their study, achieved chromatographic separation on a C8 column of 250 mm length, 4.6 mm diameter, and 5 μm particle size. The analyte was examined with a UV detector at 298 nm with a flow rate of 1 ml/min. The study conducted by Shah and Jat (16), 2015 used a mobile phase comprising buffer and methanol, pH 6.8 in the ratio of 90:10 (v/v). The eluent was analyzed at 260 nm, with a flow rate of 1 ml/min, and a run time of 20 minutes.
While these methods demonstrated acceptable validation parameters, they employed complex or buffer-based mobile phases, longer runtimes, and did not include real-world assessment of marketed formulations.
The ceftriaxone monograph in the Indian Pharmacopoeia specifies that a ceftriaxone injection contains not less than 90.0% and not more than 115.0% of the stated amount of ceftriaxone (17).
Looking at the increasing incidences of substandard drugs in the local market of Jammu and Kashmir, we have chosen ceftriaxone, which is used extensively at all levels of health care in the valley and the country, and has also been found to be “out of specification” limits on multiple occasions by the Drug and Food Control Organization of Jammu and Kashmir (18), (19), (20), (21), (22), to verify quantity and quality by employing reverse-phase HPLC (RP-HPLC)-based analytical technique.
Only a few analytical methods were reported for pharmaceutical preparations alone (15), (16), (23). Taking into consideration already reported methods, we aimed to improve upon validated methods for the estimation of ceftriaxone and develop a simple, precise, less time-consuming, and easily reproducible method which would have a wide linear range and good sensitivity for assay of ceftriaxone in the powder for injectable dosage forms.
The HPLC 1260 Infinity Model from Agilent Technologies was used for analysis of pharmaceutical formulations. The column was analytical column C18 100/4.6 mm, 3.5 μm particle size, and 20 μL sample loop size. We used EZChromS1 software in our laboratory for HPLC data processing.
The reference sample ceftriaxone standard (pharmaceutical reference standard of 99.9% purity) was obtained from Sigma-Aldrich, Mumbai, India. ACN and HPLC-grade water was procured from Merck SpecialtiesⓇ Private Limited, Mumbai, India.
Pure reference standard of ceftriaxone was used as the external standard in the analysis. Accurately weighed 50 mg of ceftriaxone was transferred to a 50 mL volumetric flask. Initially, 5 mL of the solvent system (ACN: Water; 20:80%, v/v) was added to the flask and placed on the vortex to ensure proper dissolution. This was then diluted up to the 50 mL mark to obtain a final concentration of ceftriaxone standard of 1,000 μg/mL as standard stock solution.
Standard calibration solutions were prepared from the stock solution (20 μg/mL, 30 μg/mL, 40 μg/mL, 50 μg/mL, 60 μg/mL, and 70 μg/mL) and a 6-point calibration curve was constructed.
The injectable formulation of ceftriaxone from 19 different brands was procured in triplicates from pharmacies located in urban and rural areas of different districts of Kashmir, India. Brands of ceftriaxone included in our study were numerically coded to obscure name and batch number to maintain anonymity of the brands (code number of brands tested: CEF A1-CEF A19). Stock solutions for the ceftriaxone formulations were prepared having concentration of 1,000 μg/mL. From the stock solutions, a dilution of 50 μg/mL was prepared in triplicate for estimation.
The isocratic solvent system was prepared by adding 200 mL of HPLC-grade ACN to 800 mL of HPLC-grade water in a 1,000 mL volumetric flask in a ratio of 20:80% v/v.
Initially, various mobile phase compositions of ACN and water with 1% acetic acid (ratios of 20:80, 25:75, and 30:70) were tested, but they resulted in undesirable changes in peak symmetry, finning, and retention time.
The RP-HPLC method for ceftriaxone assay was validated in terms of accuracy, precision, reproducibility, linearity, specificity, limit of detection (LOD), limit of quantification (LOQ), and robustness according to ICH Harmonized Tripartite Guidelines. To evaluate accuracy of the developed method, successive analysis (n = 3) of three different concentrations (20 μg/mL, 40 μg/mL, and 60 μg/mL) of standard ceftriaxone solution were performed at different times during the same day and on different days. Precision was evaluated by analyzing the standard drug at three concentration levels (20 μg/mL, 40 μg/mL, and 60 μg/mL). Each concentration was analyzed in triplicate at different times during the same day (intra-day precision - repeatability) and on different days (inter-day precision - intermediate precision). Percent recovery, standard deviation (SD), and relative standard deviations (RSDs) were calculated. Standard curves were constructed at concentrations 20, 30, 40, 50, 60, and 70 μg/mL of ceftriaxone.
The stability of the analytical solutions was determined by analyzing the ceftriaxone standard solution immediately after preparation. The prepared solution was stored at a temperature of 2°C-8°C and analyzed 7 days apart. Percentage (%) RSD values were calculated using a specification limit of ±2% from the initial value.
The liquid chromatographic method was isocratic in nature, employing a mobile phase mixture of ACN and water in a ratio of 20:80% v/v. The flow rate of our mobile phase was set at 0.5 mL/min and UV detection was monitored at a wavelength of 254 nm. The injection volume was 20 μL with a run time of 5 minutes.
Validation of the optimized method was performed in accordance with ICH Q2 (R1) guidelines. The validation characteristics that were addressed included specificity, LOD, LOQ, linearity and range, precision and accuracy, robustness, and solution stability.
Specificity of the RP-HPLC method was assessed by checking that no interference peaks were found at the retention time of ceftriaxone with mobile phase blank, standard and sample solutions. For this, chromatograms of solutions of standard (50 μg/mL), sample (50 μg/mL), and mobile phase blanks were compared (Figure 1).
LOD is the lowest concentration in a sample that can be detected, but not necessarily quantified under the stated experimental conditions. LOQ is the lowest concentration of analyte that can be determined with acceptable precision and accuracy. The LOD (3.3 SD/S) and LOQ (10 SD/S) were then calculated on the basis of SD of response and slope (S) of the calibration curve. An LOD value of 5.88 μg/mL was obtained. The LOQ value in our developed method was 17.83 μg/mL.
Standard curves were constructed at concentrations 20, 30, 40, 50, 60, and 70 μg/mL of ceftriaxone. A linear correlation was found between absorbance at λmax (254 nm) and concentrations of ceftriaxone in the range of 20-70 μg/mL. Curves were obtained by plotting the peak area against concentrations of ceftriaxone. Linear calibration curves were generated by linear regression analysis and obtained over the respective standard concentration ranges. The suitability of the calibration models was confirmed by back-calculating the concentrations of the calibration standards. The regression statistics are shown in Table 1.
Table 1. Linearity Data of Developed Method.
| Parameters | Values |
|---|---|
| λ max (nm) | 254 |
| Linearity range (μg/ml) | 20-70 |
| Percent recovery | 100.1 |
| Standard deviation of % recovery | 1.71 |
| %RSD | 1.7 |
| LOD (μg/mL) | 5.88 |
| LOQ (μg/mL) | 17.83 |
| Slope | 212,948 |
| Standard error on slope | 3,222.16 |
| Confidence limit of slope (95%) | 204,002.25-221,894.55 |
| Intercept | −1,372,394 (−1E + 06) |
| Standard error on intercept | 155,088.27 |
| Confidence limit of intercept (95%) | −1,802,988.36 to −941,800.22 |
| Coefficient of determination (R2) | 0.9991 |
| Λ max: detection wavelength; %RSD: percent relative standard deviation; LOD: limit of detection; LOQ: limit of quantitation. Table 1 contains the linearity data of our developed method obtained by applying linear regression statistics over the concentration range of 20-70 μg/mL. |
|
The precision of the developed method was evaluated by analyzing the pure standard drug at three different concentration levels within the linear range (20 μg/mL, 40 μg/mL, and 60 μg/mL). Each concentration was analyzed in triplicates at different times during the same day (intra-day precision - repeatability) and on different days (inter-day precision - intermediate precision). Percent recovery, SD, and RSDs were calculated. The results obtained were tabulated along with standard error and 95% confidence interval. Precision data for intra-day assay (repeatability) showed %RSD of 1.59%, while data for inter-day assay (intermediate precision) showed %RSD of 1.16%. (Table 2)
Table 2. Validation Parameters of Ceftriaxone by RP-HPLC.
| Parameters | RP-HPLC method | Recommended limits | ||
|---|---|---|---|---|
| Specificity | Chromatograms enclosed | No interferences | ||
| Accuracy, % | 99.4% | 95.0%-105.0% | ||
| Precision | Intra-day Inter-day |
1.59 1.16 |
NMT 2.00 %RSD | |
| Robustness | Flow rate (mL/min) | 0.4 0.6 |
0.96 0.48 |
NMT 2.00 %RSD |
| %ACN | 15 25 |
0.39 1.00 |
||
| Wavelength (nm) | 250 258 |
0.06 0.40 |
||
| Column temperature | 30°C 40°C |
0.87 0.40 |
||
| %RSD: percent relative standard deviation; NMT: not more than. Table 2 depicts the parameters evaluated in the validation of our developed method. For specificity of the method, relevant chromatograms have been enclosed (Figure 2). Mean accuracy of the method is 99.45% and %RSD for precision studies and robustness have values <2.00. |
||||
Accuracy is obtained by calculating the percent recovery of the analyte recovered. To evaluate the accuracy of the developed method, successive analysis (n = 3) of three different concentrations (20 μg/mL, 40 μg/mL, and 60 μg/mL), within the linear range of standard ceftriaxone solution was performed at different times during the same day and on different days. The data of the experiment was statistically analyzed to study the recovery and validity of the developed method. The mean recovery should be within 90.0%-115.0% to be accepted (Table 2).
The robustness of the method was studied by deliberate changes in the method like alterations in flow rate (0.4, 0.5, 0.6 mL/min), alteration in percent of ACN in the mobile phase (15%, 20%, 25%), changes in the wavelength (250, 254, 258 nm), and changes in column temperature (30°C, 35°C, 40°C). Representative chromatograms obtained from the robustness studies have been provided in the Supplementary Material. Data acquired from these minor variations showed %RSD values of <2% in all parameters.
The stability of the analytical solutions was determined by analyzing the ceftriaxone standard solution, stored at a temperature of 2°C-8°C 7 days apart. %RSD values of the results were calculated with the specification limit that final assay value should be within 2% of the initial value.
A consolidated summary of the validation results is presented in Table 2, while detailed data has been provided in the Supplementary Material. As shown in Table 2, the method exhibited good specificity with no interference from excipients, and the mean accuracy was found to be 99.4%, which is within the acceptable range of 95.0%-105.0%. Precision, both intra-day and inter-day, demonstrated %RSD values <2.00, indicating the reproducibility of the method. The method also proved robust under deliberate variations in flow rate, mobile phase composition, detection wavelength, and column temperature, with all %RSD values remaining well below the 2.00 threshold.
These results confirm that the developed method is reliable and suitable for routine quantitative analysis of ceftriaxone in pharmaceutical formulations.
For quantitative and qualitative analysis of ceftriaxone in pharmaceutical dosage form, we injected equal volumes of the standard and test preparation, recorded the chromatograms, and evaluated the responses for the peak of interest. Out of the 19 ceftriaxone formulations, 89.5% of the formulations (n = 17) were within the range of the Indian Pharmacopoeia specification, that is 90.0%-115.0%, while 10.5% (n = 2) of the formulations were found to be out of specification limits (Figure 2).
At expected concentration of 50 μg/ml, all the formulations contained ceftriaxone as per the standards of the Indian Pharmacopoeia (90.0%-115.0%), except for two formulations, CEF A17 and CEF A19. CEF A17 (mean percent recovery 79.6%) and CEF A19 (mean percent recovery 85.1%) were found to contain ceftriaxone below the lower permissible limit of ceftriaxone (Figure 3 and 4). At an expected concentration of 50 μg/mL of ceftriaxone, p value is statistically significant (p < 0.05). This indicates that there is statistically significant difference between the obtained mean concentrations of ceftriaxone in different formulations (Table 3). At an expected concentration of 50 μg/mL, the degree of dispersion (spread) is at the minimum in CEF A6 and CEF A11, and at the maximum in CEF A17 (Figure 5).
Table 3. Multiple Comparisons - Bonferroni.
| Dependent variable: mean | |||||||
|---|---|---|---|---|---|---|---|
| Expected concentration (CEF) | Mean difference (I-J) | Std. error | Sig. | 95% Confidence interval | |||
| Lower bound | Upper bound | ||||||
| 50 | Ceftriaxone standard | CEF A1 | −5.54667 | 1.66941 | 0.364 | −12.2301 | 1.1367 |
| CEF A2 | 1.89000 | 1.66941 | 1.000 | −4.7934 | 8.5734 | ||
| CEF A3 | 1.94667 | 1.66941 | 1.000 | −4.7367 | 8.6301 | ||
| CEF A4 | 1.94667 | 1.66941 | 1.000 | −4.7367 | 8.6301 | ||
| CEF A5 | 1.94667 | 1.66941 | 1.000 | −4.7367 | 8.6301 | ||
| CEF A6 | −4.41333 | 1.66941 | 1.000 | −11.0967 | 2.2701 | ||
| CEF A7 | −2.70333 | 1.66941 | 1.000 | −9.3867 | 3.9801 | ||
| CEF A8 | −1.72333 | 1.66941 | 1.000 | −8.4067 | 4.9601 | ||
| CEF A9 | −3.19667 | 1.66941 | 1.000 | −9.8801 | 3.4867 | ||
| CEF A10 | −3.11333 | 1.66941 | 1.000 | −9.7967 | 3.5701 | ||
| CEF A11 | −0.04333 | 1.66941 | 1.000 | −6.7267 | 6.6401 | ||
| CEF A12 | 3.69667 | 1.66941 | 1.000 | −2.9867 | 10.3801 | ||
| CEF A13 | 0.04000 | 1.66941 | 1.000 | −6.6434 | 6.7234 | ||
| CEF A14 | 3.36000 | 1.66941 | 1.000 | −3.3234 | 10.0434 | ||
| CEF A15 | 0.44000 | 1.66941 | 1.000 | −6.2434 | 7.1234 | ||
| CEF A16 | 11.17667 | 1.66941 | 1.000 | −5.5067 | 7.8601 | ||
| CEF A17 | 10.39667* | 1.66941 | 0.000 | 3.7133 | 17.0801 | ||
| CEF A18 | 1.71667 | 1.66941 | 1.000 | −4.9667 | 8.4001 | ||
| CEF A19 | 7.62667* | 1.66941 | 0.009 | 0.9433 | 14.3101 | ||
| *The mean difference is significant at the 0.05 level. In Table 3, it is clear that by comparing the ceftriaxone standard with all 19 ceftriaxone formulations at expected concentration of 50 μg/mL of ceftriaxone, there is a statistically insignificant mean difference between the different formulations, except in case of following, where the mean difference is statistically significant; CEF standard and CEF A17 (p < 0.05), CEF standard and CEF A19 (p < 0.05). |
|||||||
The developed RP-HPLC method for ceftriaxone was optimized using a relatively straightforward ACN-water system without the need for buffer or additional modifiers. This enabled us to achieve a simplified preparation, which would significantly improve reproducibility. Further, this study bridges the gap between analytical validation and public health by applying a validated method to assess market samples, revealing compliance failures in real-world settings.
One of the key challenges encountered during method development was the optimization of the mobile phase. Initially, various compositions of ACN and water with 1% acetic acid were evaluated in ratios of 20:80, 25:75, and 30:70. However, these combinations led to undesirable chromatographic behaviors such as poor peak symmetry, finning, broadened peaks, and inconsistent retention times. These anomalies significantly hindered the reproducibility and clarity of the chromatograms, making them unsuitable for reliable quantification. Ultimately, the removal of acetic acid from the mobile phase and adoption of a simpler ACN:water system in a 20:80 ratio resulted in well-resolved, symmetric peaks with stable retention times and improved reproducibility―thus forming the basis of the finalized isocratic method. Our validated method provided comparable accuracy and precision, while working at a shorter runtime (5 minutes), had a simpler mobile phase, with robustness and stability within acceptable limits.
The validation results confirmed that the method is specific, sensitive (LOD = 5.88 μg/mL), accurate (99.4% mean recovery), precise (%RSD < 2%), robust, and stable under refrigerated storage.
The results of analysis of commercial formulations agrees with findings observed in a WHO report released in 2017, wherein 10.5% of medicines sold in LMIC countries, including India, were found to be substandard and falsified (6). Similar findings were also published in a report released by the Ministry of Health and Family Welfare, Government of India in 2003, wherein it was estimated that counterfeit drugs account for 0.34% of the total pharmaceutical market and substandard drugs account for 9.34% (results varied from 8.19% to 10.64%) of the drug market in India (24).
In a study conducted by Ali, 2009 (25), out of 96 samples of ceftriaxone (of strengths 250, 500, and 1,000 mg), belonging to 33 brands, 15.62% samples (n = 15) were found out of specification range (i.e., 90.0%-115.0%). A study conducted by Sotade et al. (26), 2017 estimated an average percent recovery of 13 formulations of ceftriaxone to be between 100.3% and 103.8%. From their assay, it was concluded that the content of active ingredients was within the specified pharmacopoeial range. In a study conducted by Shah and Jat (16), 2015, the estimated average recovery percent of ceftriaxone was between 99.4% and 99.6% (n = 3). Shah et al. (14), 2013, in their study, obtained an average recovery percent of ceftriaxone of between 96.9% and 102.1% (n = 2). Arnet et al. (27), 2014 tested eight generic ceftriaxone products manufactured in Eastern Asia and found that all eight preparations failed to meet in vitro quality standards when compared to the original branded product.
Despite concerns regarding the perceived routine nature of the work, the novelty of our study lies in the application of a validated analytical method to real-world market surveillance in a region (Jammu and Kashmir) known for recurrent quality issues in antibiotics. By analyzing 19 marketed brands and identifying two as substandard, our work underscores the public health importance of simple, robust, and reproducible quality control tools. Unlike purely analytical studies focused on method development, our study integrates method validation with post-marketing surveillance, providing regulatory insights and demonstrating continued risk of substandard formulations in the pharmaceutical supply chain.
The detection of substandard products (CEF A17 and CEF A19) underscores the need for stringent quality control and post-marketing surveillance to ensure patient safety.
While the method showed acceptable validation characteristics, it was optimized for ceftriaxone alone and not designed for detection of degradation products or related impurities. The study was limited to a single geographical area (Jammu and Kashmir), and broader market surveillance may be required to generalize the findings. Matrix effects from excipients in complex formulations were not explored.
A robust, cost-effective RP-HPLC method for ceftriaxone estimation in pharmaceutical dosage form was developed and validated as per ICH guidelines. The method showed high specificity, precision, and robustness. Application to 19 brands of ceftriaxone revealed that 10.5% failed to meet Indian Pharmacopoeial standards, highlighting the need for improved regulatory oversight and routine quality control.
Younis Rather contributed to the conception and design of this study. Younis Rather and Hibba Dar performed the analysis of the pharmaceutical formulations. Hibba Dar did the statistical analyses and data management, and drafted the manuscript. Sameena Farhat and Rehana Tabassum contributed to the concept of this study and gave advice on the interpretation of the study results. Younis Rather and Rehana Tabassum reviewed the manuscript and supervised the whole study process. All authors read and approved the final manuscript.
None
This study was approved by the Institutional Ethical Committee of Government Medical College, Srinagar with Approval code IRBGMC-SGR/Pharma/913.
Universal health coverage (UHC) [Internet]. World Health Organization. [cited 2021 Nov 25]. Available from: https://www.who.int/news-room/fact-sheets/detail/universal-health-coverage-(uhc)
Nations U 70/1. Transforming our world: the 2030 Agenda for Sustainable Development Transforming our world: the 2030 Agenda for Sustainable Development Preamble. 2030. https://sdgs.un.org/2030agenda [cited 2021 Nov 25]
Howson P. Special report: India COVID-19 briefing [Internet]. Google Books. [cited 2021 Nov 25]. Available from: https://books.google.co.in/books?id=sFD5DwAAQBAJ&redir_esc=y
Indian pharmaceuticals industry analysis presentation [Internet]. IBEF. [cited 2021 Nov 25]. Available from: https://www.ibef.org/industry/indian-pharmaceuticals-industry-analysis-presentation
Kelesidis T, Falagas ME. Substandard/counterfeit antimicrobial drugs. Clin Microbiol Rev. 2015;28(2):443-64.
Executive summary [Internet]. World Health Organization. 2017 [cited 2021 Dec 15]. Available from: https://www.who.int/medicines/regulation/ssffc/publications/SEstudy-executive-summary-EN.pdf
Elander RP. Industrial production of beta-lactam antibiotics. Appl Microbiol Biotechnol [Internet]. 2003 [cited 2021 Dec 15];61(5-6):385-92. Available from: https://pubmed.ncbi.nlm.nih.gov/12679848/
Delepierre A, Gayot A, Carpentier A. Update on counterfeit antibiotics worldwide; public health risks. Med Mal Infect [Internet]. 2012 [cited 2021 Dec 15];42(6):247-55. Available from: https://pubmed.ncbi.nlm.nih.gov/22621827/
WHO global surveillance and monitoring system for substandard and falsified medical products [Internet]. World Health Organization. 2017 [cited 2021 Dec 15]. Available from: https://apps.who.int/iris/handle/10665/326708
Pharmaceutical analysis in quality control [Internet]. European Pharmaceutical Review. 2025 [cited 2022 Jan 06]. Available from: https://www.europeanpharmaceuticalreview.com/article/868/pharmaceutical-analysis-in-quality-control/
Mennickent S, de Diego M. Analytical method validation as the first step in drug quality control. In: Pereira P, Xavier S, editors. Quality management and quality control - new trends and developments [Internet]. IntechOpen; 2019 [cited 2022 Jan 06]. Available from: https://www.intechopen.com/chapters/65518
Siddiqui MR, AlOthman ZA, Rahman N. Analytical techniques in pharmaceutical analysis: a review. Arab J Chem. 2017;10:S1409-21.
Ahmed S, Islam MS, Ullah B, et al. A review article on pharmaceutical analysis of pharmaceutical industry according to pharmacopoeias. Orient J Chem. 2020;36(1):1-10.
Shah J, Rasul Jan M, Shah S, et al. Development and validation of HPLC method for simultaneous determination of ceftriaxone and cefaclor in commercial formulations and biological samples. J Mex Chem Soc [Internet]; 2013 [cited 2022 Jan 07];57(4):314-20. Available from: http://www.scielo.org.mx/scielo.php?script=sci_arttext&pid=S1870-249X2013000400008&lng=es&nrm=iso&tlng=en
Gurupadayya BM, Disha NS. Stability indicating HPLC method for the simultaneous determination of ceftriaxone and vancomycin in pharmaceutical formulation. J Chromat Separation Techniq. 2013;4(10):2-5.
Shah R, Jat R. Development and validation of reversed phase HPLC method for estimation of ceftriaxone in pharmaceutical dosage form. J Drug Delivery Ther. 2015;5(3):87-91.
Indian pharmacopoeia 2014. The Indian Pharmacopoeia Commission, Central Indian Pharmacopoeial Laboratory, Govt of India, Ministry of Health and Family Welfare.
List of sub-standard drugs in Jammu & Kashmir state during the period of 2016-17; 2016. https://dfcojk.org/pdf/NSQ%20upto%20February%202018%20%20(2017-18).pdf [cited 2022 Jan 19]
List of sub-standard drugs in Jammu & Kashmir state during the period of 2015-16; 2015.
List of sub-standard drugs in Jammu & Kashmir state during the period of 2013-14. https://dfcojk.org/pdf/sub_2013.pdf [cited 2022 Jan 19]
List of sub-standard drugs in Jammu & Kashmir state during the period of 2012-13. https://dfcojk.org/pdf/sub_2012.pdf [cited 2022 Jan 19]
List of sub-standard drugs in Jammu & Kashmir state during the period of 2017-18. https://dfcojk.org/pdf/NSQ%202017-18.pdf [cited 2022 Jan 19]
Shrivastava SM, Singh R, Tariq A, et al. A novel high performance liquid chromatographic method for simultaneous determination of ceftriaxone and sulbactam in Sulbactomax. Int J Biomed Sci. 2009;5(1):37-43.
Report of the expert committee on a comprehensive examination of drug regulatory issues, including the problem of spurious drugs [Internet]. Government of India: Ministry of Health and Family Welfare. 2003 [cited 2022 Jan 19]. Available from: http://cdsco.nic.in/html/Final20Report20mashelkar.pdf
Obaid A. Quality of ceftriaxone in Pakistan: reality and resonance. Pak J Pharm Sci. 2009;22(2):220-9.
Sotade AA, Idowu PA, Adegoke OA. Comparative quality assessment of generic brands of ceftriaxone sodium injection marketed in Ibadan, South- West Nigeria. J Pharm Bio. 2017;14(2):175-86.
Arnet I, Altermatt M, Roggo Y, et al. Pharmaceutical quality of eight generics of ceftriaxone preparation for injection in Eastern Asia. J Chemother [Internet]. 2015 [cited 2022 Jan 19];27(6):337-42. Available from: https://www.researchgate.net/publication/264989225_Pharmaceutical_quality_of_eight_generics_of_ceftriaxone_preparation_for_injection_in_Eastern_Asia